CW Write absolute counter position

Command Command description Controller response Default
mode
(N/A - not available)
Service
Mode
(N/A - not available)
xCWy Write the absolute counter position. y: positive, negative, or zero position. 1 micro step = 1/8 of motor step - Yes Yes

SYNTAX    xCWy, where x - controller (axis) number.
PARAMETER y [number] 1 micro step = 1/8 of stepper motor step
Range -2000000000...2000000000.
DESCRIPTION: Write the absolute counter position. Presets internal steps counter value with the parameter y. The counter value can be read by CR command.
RETURNS  none.
REL. COMMANDS CR, EN, EZ,
EXAMPLES: 5 CW 0   ||  3cw-500   ||  2345cW 20